top of page

CO2 Incubator C16

Advantages
REF 14209 (without O₂ control)
REF 14452 (with O₂ control)

Incubator with an inner volume of 16 liters
CO2-sensor:
  • Dual beam infrared sensor

  • Very precise measurement

  • Very short recovery times

  • Long-term stable

Direct Heating:
  • Independent over-all heating of inner walls and door

  • Homogeneous temperature distribution in the interior

  • Very short recovery times

Optional:
  • O2-control

  • PC software LaboDat+ 

Very compact and space saving construction

pro2.png
Texhnical Data:
Exterior:

Interior:

Temperature:
CO₂-control:
O₂-control < 21 % (optional):
Supply:
Supply:
Dimensions 290 x 330 x 480 mm (w x d x h)
Volume 16 l
Electro-polished stainless steel
2 shelves 205 x 230 mm (w x d)
Panel heated interior chamber and door
Range 27 °C - 42 °C starting 5 °C above ambient temperature
Stability / uniformity: ± 0.1 °C / ± 0.3 °C
Dual beam infrared sensor
Range 0 – 10 % CO₂
Stability ± 0.1 % CO₂
Measurement with galvanic sensor
Connection of N₂
Range 1 % – 21 % O₂
110 – 230 V AC, 50/60 Hz, 60 W
Inlet pressure for gas 0.8 bar
Ambient temperature 18 °C - 30 °C

Safety class I
Class IIa for every usage according to EC-Directive 93/42/EEC

EN 61010 conform

bottom of page