InControl 1050
Advantages
Features:
-
Accurate measuring of CO2, temperature and optional O2
-
Single or continuous measurement in arbitrary intervals
-
Measurement and documentation of multiple incubators
-
Documentation of measurements with date, time and incubatornumber
-
Easy menu navigation
-
Data-Download via USB; PC software DataVISUAL`09 incl.
-
Continuous operation by Li-Ion battery or mains adapter possible.
Optional accessories:
-
REF: 10556
Sensor for O2 measurement
-
REF: 15064
Flat tape sensor PT 1000 for temperature measurement inside Labo C-Top or other benchtop incubators
-
REF: 10537
Surface sensor for temperature measurement


Technical Data:
Case:
-
Plastic (ABS)
CO2 measurement:
-
Dual beam IR sensor
-
Range: 0 - 10% CO2
-
Display definition: 0.1% CO2
-
Accuracy:
- 0 - 6% CO2 ± 0.2% CO2
- 6 - 10% CO2 ± 0.3% CO2 -
Sample volume per
measurement (2 min.):
approx. 0.8 l
Temperature measurement:
-
Platinum-temperature sensor PT1000
-
Range: 0 - 100°C
-
Display definition: 0.1°C
-
Accuracy:
- 20 - 50°C ± 0.2°C
- 50 - 100°C ± 0.3°C
O2 measurement:
-
Galvanic sensor
-
Range: 0 - 100% O2
-
Display definition: 0.1% O2
-
O2-sensor not included
Data logger:
-
Recording of measurement logging in arbitrary intervals
-
Output of measurements via:
- backlit LCD
- PC software Data Visual’09
(via USB data cable) -
Interval for data logging: 15-120 min.
adjustable in steps of 5 minutes -
Storage of max. 1008 measurements
with date and time
Power supply:
-
Rechargeable Li-Ion battery
-
Power supply 100 – 240 V AC, 5 V DC
1.5 A
Operation time:
-
Use of Rechargeable battery approx.
4 hrs -
Time to recharge battery approx. 4 hrs