top of page

CO₂ Incubator C60

Incubator with an inner volume of 60 liters

  • REF 14093

  • Approved medical device

  • Active sterile humidity:

    • External water reservoir

    • Distilled water is vaporized at 120 °C

    • Minimized risk of contamination

    • Humidity sensor (rH)

    • Very fast recovery

  • CO2 sensor:

    • Dual beam infrared sensor

    • Very precise measurement

    • Very fast recovery

    • Long-term stable

  • Direct heating:

    • Independent six-sided direct heating

    • Homogeneous temperature distribution in incubation chamber

    • Very fast recovery

  • Optional:

    • O2-control

    • Subdivided inner glass door with 4 small doors

    • Lockable front door

    • Access port

    • PC software LaboDat+

CO₂-Incubator-C60.png
Technical Data:
Exterior:
Interior:
Temperature:
  • Dimensions 510 x 540 x 720 mm (w x d x h)

  • Chamber volume 60 l

  • Electro-polished stainless steel

  • Shelves height adjustable

  • Panel heated interior chamber and door

  • Range 27-42 °C starting 5 °C above ambient temperature

  • Stability / uniformity ±0.1 °C / ±0.3 °C

  • Over-temperature alarm 32-47 °C

CO2 control:
  • Dual beam infrared sensor

  • Range 0–10 % CO2

  • Stability ±0.1 % CO2

O2 control < 21 % (optional)
Humidity:
  • Measurement with galvanic sensor

  • Connection of N2

  • Adjustable range 1–21 % O2

  • Active sterile humidity generated by vaporizing module at 120 °C
  • Measuring range 0–98 % rH / range 60–95 % rH
Supply:
  • 230 V AC, 50/60 Hz or 115 V AC, 50/60 Hz, 160
  • Inlet pressure for gas 0.8 bar
  • Ambient temperature 18-30 °C
Classification:

Safety class I

Class IIa for all usage according to EC-Directive 93/42/EE

Conform to ENN 61010

bottom of page