top of page

Cell-Trans 4016

Advantages
Features:
  • Secure transport of cells without electric mains and CO2 connection

  • Factory-aligned parameters: 37°C and 5% CO2

  • Different temperature or CO2 concentration possible (to be advised before order)

  • Operation at 12V car voltage possible

  • Use of different heating blocks

  • Easy handling and cleaning

  • Compact Design 

lab7.png
Technical data:
Power supply:
Dimensions:
  • Input: 100 – 240 V AC, 50/60 Hz

  • Output: 12V DC power supply or with car voltage via adapter cable

  • Outside: 320 x 190 x 280 (l x w x h in mm)

  • Inside: 210 x 140 x 170 (l x w x h in mm)

Weight:
  • 2.4 kgs

Temperature:
  • Available between 32°C and 39°C (to be advised before order)

Heating up time:
  • approx. 2 hrs.

CO2 concentration:
  • 5% (different concentrations possible)

  • Heating block inside necessary

  • Connection to CO2 supply with -pre-pressure of 0.8 bar necessary

Parameters during transport without connection to electric mains and CO2 bottle:
Transport time:
  • 0 min

  • 10 min

  • 30 min

  • 60 min

Temperature:

  • 37°C

  • 36.5°C

  • 35.3°C

  • 34.5°C

CO2 concentration:

  • 5%

  • 4.9%

  • 4.8%

  • 4.5%

bottom of page