top of page

Aspirator 3

Advantages
Precise aspiration pump for follicular puncture:
  • REF: 14193 (without heating stage)

  • REF: 13907 (with heating stage)

Features:
  • Approved medical device

  • Easy handling by electrical pedal

  • Long durability

  • High reliability

  • Very silent

  • Constant control of negative pressure

  • Safe by overflow protection

Options:
  • integrated heating stage

  • movable support stand

lab91.png
lab92.png
Technical Data:
Power supply:
Dimensions:
  • 230 V AC, 50/60 Hz or 115 V AC,

  • 50/60 Hz

  • 80 W

  • 408 x 130 x 250 (w x d x h in mm)

Hotplate (optional):
  • Temperature range: 30°C - 45°C 

  • Accuracy: +/-0.2°C 

  • Temperature distribution: +/-0.5°C

  • Heating up time: <10 mins

According to Directive 93/42/EEC:
  • Class IIa

EN 10079 conform:
Safety class:
Device type :
  •  high vacuum

  • low flow

I
BF
Weight:
  • 7.6 kg

Negative pressure:
  • 0 - 500 mmHG ≙ 0 – 666 mbar

Volume of overflow vessel:
  • 100 ml

Connection for aspiration tubing:
  • Ø 2 to 4 mm

Connection for potential compensation